204 Technology Park Lane
Fuquay-Varina,
North Carolina 27526
Toll Free: 1-800-882-3214
International: +1 919-557-7300
Gas Analyzer for Continuous Emissions Monitoring
MultiGas™ 2030 CEM is an FTIR-based gas analyzer designed for continuous monitoring of emissions in effluent streams containing up to 40% of water. Permanently stored calibration spectra reduces the need for costly reference gas mixtures.
MultiGas™ 2030 analyzer contains a 2102 Process FTIR Spectrometer, a patented, high-optical-throughput sampling cell, applications-specific analysis software, and an instrument-independent, quantitative spectral library. It incorporates a novel long wavelength thermoelectrically (TE) cooled detector for more accurate, highly sensitive measurement of most gases and vapors in high moisture streams (up to 40%), producing high resolution spectra (0.5cm-1), enabling detection and measurement of components such as SO2, NO2 and NO, without removing the
moisture.
Applications:
Cement Kilns
Waste Incinerators
Power Plants
Large Combustion Plants
Turbine Engines
Analyzer
Measurement Technique FTIR Spectrometry
Spectral Resolution 0.5 cm-1
Scan Time 60 seconds
Infrared Source Silicon Carbide @ 1200°C
Reference Laser Helium Neon (15798.2 cm-1)
Detector Thermoelectrically (TE) cooled MCT (HgCdTe)
Purge Pressure 20 psig (1.5 bar) max.
Spectrometer Purge Flow 0.2 L/min of dry nitrogen or CO2 free clean dry air
with dewpoints below -70°C
Optics Purge Flow 0.2 L/min of dry nitrogen or CO2 free clean dry air
with dewpoints below -70°C
Pressure Transducer MKS Baratron® Capacitance Manometer
Purge Connection 1/4 inch Swagelok® quick connect
Dimensions 17.5"W x 12.5"H x 25.5"D
Installation 19" rack mount chassis
Power 120 or 240 VAC, 50/60 Hz, 3 amps
Weight 110 lbs. (50 kg)
Sample Parameters
Sample Temperature 191°C
Sample Flow 1 L/min
Sample Pressure 1 atm ±0.05
Gas Cell
Construction Nickel coated Al
Fittings 1/4 inch Swagelok®
Tubing Heated 1/4 inch stainless steel
Mirrors Nickel plated aluminum substrate,
with rugged gold coating
Windows BaF2
O-rings Viton®
Computer Requirements/Communication Options
Computer Requirements Intel Pentium® PC, Microsoft®
Windows® XP or Windows® 7
operating systems, display (1024 x 768)
Desktop or Notebook
Minimum Recommended Computer Intel Pentium® III, 850 MHz,
Microsoft® Windows® XP, 256 MB
Computer/FTIR Communications RJ-45 Crossover Ethernet
Communication Protocol TOOLweb® (HTML based), OPC,
Modbus TCP/IP