Click on image to view or download.

|
The Apex Instruments 3-D Pitot is capable of measuring
yaw and pitch of nonuniform flows. The yaw angle is determined
directly by rotating the pitot to null the pressure across
a pair of symentrically placed ports on the pitot head.
The yaw angle is measured with a remote inclinometer attached
to the probe extension and digitally displayed on the console.
The pitch angle is calculated using probe specific calibration
curves. The average gas volumetric flow rate in the stack
or duct is then determined from the average axial velocity.
3-D Pitot Calibration (PT-CAL-3D)
All 3-D pitot tubes must be calibrated before use. Recalibration
is also required either within 12 months of its first field
use after its most recent calibration or after 10 field
tests, whichever occurs later. In addition, whenever there
is visible damage to the 3-D head, the pitot shall be recalibrated
before it is used again.
|