The XC-5000 AutoKintiec™ Source Sampler is designed for conducting US EPA Method 5 and associated isokinetic methods. Take the worry out of isokinetic calculations and the human error out of data entries recorded by hand. The new XC-5000 is compatible with your existing Method 5 stack sampling components. Report preparation is streamlined with accurate data downloadable to files for easy preparation and submission to your governing agency.
All applicable components are rated 110V / 60Hz.
• Uses standard Apex Instruments sample train components
(probe, nozzle, heated filter box, impinger box, etc.)
• Maintains isokinetic sampling conditions without operator assistance
• Calculates stack traverse points
• Calculates correct nozzle diameter for site conditions
• Stores and displays user entered data and parameters
• Upload and download data files
• Performs instantaneous isokinetic sample flow calculations
• Timer alerts operator to move the probe
• Alarms indicate when acceptable conditions are not met
• English or metric units selectable
Gas Meter: Model AP25, Precision DGM, 0.7 liters per revolution,
digital gas volume with Internal Quadrature Encoder with 6 digit LCD
display, 1cc resolution
Temperature Control: Integrated temperature control via the Control
and Data Acquisition Board, probe and oven with solid state relays.
Thermocouple Display: 7 temperatures displayed simultaneously on
the PC User Interface, °F or °C, Probe, Stack, Oven, Filter, Exit, AUX
and DGM
Digital Pressure Transducers for H and P Bi-Directional, Barometric
Delta P +/- 2.5” 0.01” resolution
+/- 63mm 0.01 mm resolution
Delta H 0”-5” 0.01” resolutions
0mm-127mm 0.01mm resolution
Barometric 17.7 inHg – 32.5 inHg 0.01 inHg resolution
450 mmHg – 825 mmHg 0.01m mmHg resolution
Optional:
4 channel analog input module for logging external
data (4-20ma, 0-10V, 1-5V )
Dimensions:
23” x 21” x 12” (58 cm x 53 cm x 30.5 cm)
Weight: 39 lbs. (17.7 kg).
Power: 120V / 60 Hz.
Console Power Requirements:
120V 15amp max
Umbilical Connections:
Electrical: 4 conductor circular connector grounded shell.
Sample line: Stainless Steel 1/2” Quick Connector.
Pitot Line: Stainless Steel 1/4” Quick Connectors (optional 3/8”).
External pump: Stainless Steel 3/8” Quick Connect.
Thermocouples: Type-K standard size.
Communication:
Wireless and Ethernet