Please See Alternative ALTEF and Multi-Layer Foil Gas Sampling Bags Comparison Charts.
Gas Sampling Bags are the first choice for collecting a variety of gaseous compounds. Gas Sampling Bags are referenced for many U.S. EPA Methods including 3,18 and 0040. Bags fabricated from film are inert to most gases, resist gas permeation and can be reused for most applications after proper cleaning. They are flexible and strong, and use of a rigid container ensures a longer life and convenience during transporting. Gas Sampling Bags are fabricated from transparent 2ml film or can be special ordered fabricated from opaque black for photosensitive compounds. Custom sizes available upon request. Sample bags come standard fitted with the Poly 2-in-1 Combination Valve and Septum Fitting. Each standard bag has one fitting located near the top center or in the center of the bag. Custom placement available.
| Compound | Recommended Gas Sampling Bag Material |
|
| ALTEF | Multi-Layer Foil | |
| n-Butyl mercaptan | ||
| tert-Butyl mercaptan | ||
| Carbon Disulfide• | ||
| Carbonyl Sulfide | ||
| Diethyl disulfide | ||
| Diethyl Sulfide• | ||
| Dimethyl disulfide | ||
| Dimethyl Sulfide• | ||
| 2, 5-Dimethylthiophene | ||
| Ethyl mercaptan• | ||
| Ethyl methyl Sulfide• | ||
| 2-Ethylthiopene | ||
| Hydrogen Sulfide | ||
| Isobutyl mercaptan• | ||
| Isopropyl mercaptan• | ||
| 3-Methylthiophene | ||
| Methyl mercaptan• | ||
| n-Propyl mercaptan• | ||
| Tetrahydrothiophene | ||
| Thiophene• |
| Compound | Recommended Gas Sampling Bag Material |
|
| ALTEF | Multi-Layer Foil? | |
| Acetone, Acetonitrile, Acrylonitrile, Butyl Acetate, Ethyl Acetate, Methyl Ethyl Ketone, Toluene, p-Xylene |
||
| Allyl Chloride, Benezine, Bromoethane Carbon Tetrachloride, Chloroform Carbon Dioxide, Carbon Monoxide |
||
| 1,2-Dichloroethane, Dichloropropane Ethylene, Heptane, Hexane, Isooctane Isopropyl Alcohol, Methane |
||
| Methylene Chloride, Methy tert-butyl ether Octane, Percholoethylene, Propylene Propylene Oxide, Tetrahydrofuran |
||
| 1, 1, 1-Trichloroethane, Trichloroethylene Vinylidene Chloride |
Color Code:
Red - Not Suitable
Dark Green - Recommended
Ligth Green - Suitable when used as recommended
• ALTEF bags can be used to sample these sulfur compounds if sample is analyzed within 24 hours.
◊ Multip-Layer Foil bags can be used to sample most VOCs but are not recommended for collecting low ppm to
high ppb VOCs due to background levels from bag materials.
ALTEF bags are recommended for most VOCs, if analyzed within 48 hours, and for many sulfer compounds, if
analyzed within 24 hours.
Multi-Layer Foil bags are recommended for Methane (CH4), Hydrogen Sulfide (H2S), Carbon Monoxide (CO),
and Carbon Dioxide (CO2), if analyzed within 48 hours.
| Description | Poly 2-in-1 Combo Valve | Brass Roberts Valve |
|---|---|---|
| 6" x 6" (0.6 liter) | TB-6-P | TB-6-R |
| 7” x 7” (1.0 liter) | TB-7-P | TB-7-R |
| 9” x 9” (2.0 liter) | TB-9-P | TB-9-R |
| 10” x 10” (3.0 liter) | TB-10-P | TB-10-R |
| 12” x 12” (5.0 liter) | TB-12-P | TB-12-R |
| 12” x 15” (7.0 liter) | TB-12X15-P | TB-12X15-R |
| 12” x 17” (8.0 liter) | TB-12X17-P | TB-12X17-R |
| 12” x 18” (9.5 liter) | TB-12X18-P | TB-12X18-R |
| 12” x 19” (10.0 liter) | TB-12X19-P | TB-12X19-R |
Show all Gas Sampling Bags (32)