HAVE A QUESTION? CALL US!
(800) 882-3214

HM-T96 Humidity Monitor

HM-T96 Humidity Monitor

The Apex Humidity Monitor is designed for monitoring the relative humidity of the sample between the sample pump and flow panel to protect gas analyzers.

Description

The module will alert you of dangerous levels before the dew point is reached and will shut down the sample pump to protect the analyzers. There’s also a redundant electro-optical switch for detecting condensate located in a stainless-steel coalescing filter assembly.

The HM-T96 uses a Telaire T9602 Humidity and Temperature Sensor, which is designed for harsh environments.

Humidity Monitor Specifications:

Method: capacitive polymer RH sensor, PTA (Proportional to Absolute) integrated temperature sensor
Operating conditions: -20ºC to 70ºC, 0% to 100% RH
Output mode: digital model
Cable length: up to 1.8 m or 70 inches

HM-T9602 Humidity Specifications:

Resolution: 14 bit (0.01% RH)
Accuracy:* ±2.0% RH (20~80% RH), ±3.5% RH (0% to 20%) and (80% to 100%)
Repeatability: ±0.2% RH
Hysteresis: ±1.0% RH
Linearity: <2.0% RH
Response Time:** :;; 29 sec (T 63%)
Temp Coefficient: 0.13% RH/°C (at 10~60°C, 10~90% RH)
Operating: 0 ~ 95% RH
Long Term Drift: <0.5% RH/year (Normal conditions)

* Accuracies measured at 25°C, nominal voltage.
** 30% RH step response, measured at 25°C in a 1 m/sec air flow.

HM-T9602 Temperature Specifications:

Resolution: 14 bit (0.01°C)
Accuracy:* ±0.5°C
Repeatability: ±0.1°c
Response Time:** :;;116 sec (T 63)
Operating Range:*** -20°C to 70°C
Long Term Drift: <0.05°C/year (Normal conditions)

* Accuracies measured at 25°C, nominal voltage.
** 25°C step response in a 1 m/sec air flow.
*** Minimum design range, documented testing